Lab 10T 1000C 20-segment Programmable Automatic High Temperature Heat Press Machine 



Operation interface (Standard mode)

 

Standard interface: Various parameters can be set in this interface individually: pressure, demoulding, time, sample size, etc

Pressurization button: Press according to the parameters set on the interface.

Stop button: Press the stop button during the pressurization process to stop the pressurization.

Demoulding button: Press the demoulding button to release the sample according to the demoulding pressure.

Timing switch: The timing function can be turned on and off.

 

Operation interface (Advanced mode)

 

Activate button: Can be pressed according to the multi-stage pressurization program.

Stop button: During the pressurization process, you can stop the pressurization process.

Setting button: Click the Settings button to enter the Settings menu.

 

Operation interface (Advanced mode)

 

List/Curve: The current display state can be switched between list and curve.

 

Setting interface

 

Step 1: Select the data screen and enter the Settings screen

Step 2: Select "Programmed Segment Settings"

Step 3: There are 20 temperature control procedures in the program, select the first paragraph

Step 4: According to the requirements of the experiment, set the temperature and heating speed of the first stage

Step 5: If you need multiple heating procedures, you can set up 20 procedures

Step 6: After setting, press the main screen to return to the operation screen

 

Operating interface

 

Step 7: On the main screen, press "Heating start"

Step 8: The device is programmed to heat up.

Step 9: Automatically stop the device after the end of heating. If the setting is wrong, for example, the set temperature is not reached within the specified time, the device will automatically terminate.

 

Operation procedure

 

 

 

 

 

Technical parameter

Instrument type

PP-900L

Pressure range

0-10.0 tons

Pressurization process

Programmed pressure - programmed pressure retention - timed pressure relief

Holding time

1 s to ∞ s

Cylinder stroke

80mm

Heating temperature

Up to 1000℃

Mold material

Nickel base alloy (high temperature resistant material)

Sample size

Ф 10-30 mm

Mold contour

Ф 50×90mm

Furnace aperture

Ф 60 mm

Overall size

400×380×780(L×W×H)

Power source

220V 50Hz

Hot press size diagram